Switching on and Use of Turbo-V 1K-G
The following instructions apply to the Turbo-V 1K-G used togheter
with Navigator controller (model 9698978M005). Please refer to the
relative control unit manual for use with other controller model
(9699454M002).
To switch on the Turbo-V 1K-G it is necessary to supply the mains.
The integrated controller automatically recognizes the interlock and
start signals presence and start up the pump.
The first pump start up is in “Soft Start” mode. When the start up
cycle is finished, the “Soft Start” mode automatically is disabled, and
the following start ups are without the “Soft Start” mode. To re-
enable the “Soft Start” mode it must be activated by the suitable
software command (see the paragraph “RS 232/RS 485
COMMUNICATION DESCRIPTION” in the appendix “Technical
Information”).
The green LED located on the Navigator controller base front panel
indicates with its flashing frequency the system operating conditions:
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in “Waiting for interlock” status;
fast flashing (period of about 200 ms): error condition.
Summary of Contents for 8698961R002
Page 3: ...Turbo V 1K G Turbo V 1K G User Manual 87 901 025 01 A 3 304 Turbo V 1K G...
Page 4: ...Turbo V 1K G 4 304 Turbo V 1K G User Manual 87 901 025 01 A...
Page 14: ...Contents 14 304 Turbo V 1K G User Manual 87 901 025 01 A...
Page 58: ...3 Mode d emploi Mise au Rebut 58 304 Turbo V 1K G User Manual 87 901 025 01 A...
Page 88: ...5 Manual de Istru es Elimina o 88 304 Turbo V 1K G User Manual 87 901 025 01 A...
Page 95: ...Bedrijfshandleiding Uitpakken 6 Turbo V 1K G User Manual 87 901 025 01 A 95 304...
Page 104: ...6 Bedrijfshandleiding Afvalverwerking 104 304 Turbo V 1K G User Manual 87 901 025 01 A...
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......