85056D
1- 7
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of cable
connectors and adapters is referred to in terms of the center conductor. For example, a
connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
8510-series, 872x, and 875x ONLY:
In contrast, during a measurement calibration, the
network analyzer softkey menus label a 1.85 mm calibration device with reference to the
sex of the analyzer’s test port connector—not the calibration device connector. For
example, the label
SHORT(F)
refers to the short that is to be connected to the female test
port. This will be a male short from the calibration kit.
A connector gage is referred to in terms of the connector that it measures. For instance, a
male connector gage has a female connector on the gage so that it can measure male
devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in the kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of these techniques are described in
. Failure to detect and remove dirt or
metallic particles on a mating plane surface can degrade repeatability and accuracy and
can damage any connector mated to it. Improper connections, resulting from pin depth
values being out of the observed limits (see
), or from bad connection
techniques, can also damage these devices.
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the systematic
Table 1-2 Clarifying the Terminology of Connectors: Examples
Terminology
Meaning
Short
–f–
Female short (female center conductor)
Short (f)
Male short (male center conductor) to be connected to female port
Summary of Contents for 85056D
Page 4: ...iv 85056D...
Page 6: ...vi 85056D Contents...
Page 7: ...1 1 1 General Information...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85056D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Page 38: ...4 4 85056D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Page 49: ...A 1 A Standard Definitions...