Electrostatic Discharge
Electrostatic Discharge
Protection against electrostatic discharge (ESD) is essential while connecting, inspecting,
or cleaning connectors attached to a static-sensitive circuit (such as those found in test
sets).
Static electricity can build up on your body and can easily damage sensitive internal
circuit elements when discharged. Static discharges too small to be felt can cause
permanent damage. Devices such as calibration components and devices under test (DUT),
can also carry an electrostatic charge. To prevent damage to the test set, components, and
devices:
•
Always
wear a grounded wrist strap having a 1 M
Ω
resistor in series with it when
handling components and devices or when making connections to the test set.
•
Always
use a grounded antistatic mat in front of your test equipment.
•
Always
wear a heel strap when working in an area with a conductive floor. If you are
uncertain about the conductivity of your floor, wear a heel strap.
•
Always
ground yourself before you clean, inspect, or make a connection to a
static-sensitive device or test port. You can, for example, grasp the grounded outer shell
of the test port or cable connector briefly.
•
Always
ground the center conductor of a test cable before making a connection to the
analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
shows a typical ESD protection setup using a grounded mat and wrist strap.
Refer to
for information on ordering supplies for ESD protection.
Figure 3-1 ESD Protection Setup
Summary of Contents for 85056D
Page 4: ...iv 85056D...
Page 6: ...vi 85056D Contents...
Page 7: ...1 1 1 General Information...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85056D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Page 38: ...4 4 85056D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Page 49: ...A 1 A Standard Definitions...