3-4
85056D
Use, Maintenance, and Care of the Devices
Inspecting Connectors
Inspect Female Connectors
Inspect the contact fingers in the female center conductor carefully. These can be bent or
broken, and damage to them is not always easy to see. A connector with damaged contact
fingers will not make good electrical contact and must be replaced.
NOTE
This is particularly important when mating nonprecision to precision devices.
The female 2.4 mm connectors in this calibration kit are metrology-grade, precision
slotless connectors (PSC). Precision slotless connectors are used to improve accuracy. With
PSCs on test ports and standards, the accuracy achieved when measuring at 50 dB return
loss levels is comparable to using conventional slotted connectors measuring devices
having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
Conventional female center conductors are slotted and, when mated, are flared by the
male pin. Because physical dimensions determine connector impedance, this change in
physical dimension affects electrical performance, making it very difficult to perform
precision measurements with conventional slotted female connectors.
The precision slotless connector was developed to eliminate this problem. The PSC has a
center conductor with a solid cylindrical shell, the outside diameter of which does not
change when mated. Instead, this center conductor has an internal contact that flexes to
accept the male pin.
Summary of Contents for 85056D
Page 4: ...iv 85056D...
Page 6: ...vi 85056D Contents...
Page 7: ...1 1 1 General Information...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85056D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Page 38: ...4 4 85056D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Page 49: ...A 1 A Standard Definitions...