4-5
Performance Tests
Tests for Sources
5
Compare the measured values to the specified values listed in
“Specifications and Regulatory Information”
.
2. Average Output Power
Do
not
modulate the source during this procedure.
1
Connect the equipment as shown in
.
Figure 4-2. Average output power test setup
2
Turn on the Agilent 83400-series source and an optical power meter, and allow
them to warm up for 1 hour.
3
Calibrate the optical power meter, and then measure the output power of the
source. Compare the measured value to the specified power level listed in
Chapter 3, “Specifications and Regulatory Information”
.