1-12
Getting Started
Entering Calibration Data into an Agilent 8702A/B
To install from an external disk drive
If you have an HP 9122-series dual-sided disk drive connected to your
Agilent 8702A/B, use this procedure to enter the source or receiver’s calibra-
tion data.
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word
RECEIVER
for
SOURCE
and the word
RCVR
for
SRC
.
1
Locate the LIF
-
formatted calibration disk provided with the source.
2
Place the disk in the disk drive.
3
On the Agilent 8702A/B, press:
PRESET
LOCAL
,
SYSTEM CONTROLLER
CAL
,
CAL KITS & STDS
,
SOURCE [COEFF]
,
CAL STD: SRC DISK
LOAD SRC DISK
,
READ FILE TITLES
,
LOAD <filename>
.