background image

 
 

 

 

12 

7890 Series GC Site Preparation Checklist 
 

Issued: 09-Dec-2015, Revision: 1.8 

Copyright 

 2013   

Agilent Technologies      

 

Page 12 of 15 

 
 

 

Considerations for Hydrogen Carrier Gas 

If planning to use hydrogen carrier gas, note that special considerations apply due to hydrogen’s  
flammability and chromatographic properties. Refer to the to the "Gas Supplies/Requirements for 
Hydrogen as a Carrier Gas" section in the "Agilent GC, GC/MS and ALS Site Preparation Guide" for more 
detail. 
 

Hydrogen Safety 

When using hydrogen as the carrier gas or fuel gas, be aware that hydrogen gas can flow into the GC 
oven and create an explosion hazard. Therefore, be sure that the Hydrogen gas supply is turned off until 
all connections are made and ensure the inlet and detector column fittings are either connected to a 
column or capped at all times when hydrogen gas is supplied to the instrument.  
In any application using hydrogen, leak test all connections, lines, and valves before operating the 
instrument.  
Agilent highly recommends the G3388B Leak Detector or equivalent to safely check for leaks. 
 

Supply tubing for Hydrogen Gas 

Agilent recommends using NEW, chromatographic quality copper or stainless steel tubing and fittings 
when using hydrogen. 
Do not re-use old tubing when installing or switching to hydrogen carrier gas. Hydrogen gas tends to 
remove contaminants left on old tubing by previous gases (by helium, for example).  
These contaminants can appear in detector output as high background noise or hydrocarbon 
contamination for several weeks. 
Do not use old copper tubing with hydrogen gas. Old copper tubing can become brittle and create a 
safety hazard. 
 

Hydrogen Gas Supplies 

Hydrogen can be supplied from a gas generator or from a cylinder.  

Agilent recommends use of a high-quality hydrogen gas generator. A high-quality generator can 
consistently produce purity > 99.9999%, and the generator can include built-in safety features such as 
limited flow rates, and auto-shutdown.  

If using a hydrogen gas cylinder, Agilent recommends use of Gas Clean Filters to purify the gas.  

Consider additional safety equipment as recommended by your company safety personnel. 

 

 

 

Summary of Contents for 7890 GC

Page 1: ...ns for the lab and gas venting Laboratory gases and plumbing The power requirements related to the product The required operating supplies necessary for the product at installation Please consult the Other Requirements section for other product specific information 6BIf Agilent is delivering installation and familiarization services users of the instrument should be present throughout these servic...

Page 2: ... includes a GC and a computer requires about 86 cm of bench space 2 Avoid bench space with overhanging shelves A 7683 or 7693 automatic liquid sampler will add to the height of the instrument as shown below 3 G1888A Headspace 5977 GCMS and QQQ MS are installed to the left of the 7890 and the 7697 Headspace and 220 240 Ion Trap MS are installed to the right Refer to the Dimensions and Weight sectio...

Page 3: ...table for optimum performance 3 For storage or shipping the allowable temperature range is 40 to 70 C and the allowable humidity range is 5 95 non condensing After exposing the GC to extremes of temperature or humidity allow 2 hours for it to return to the recommended ranges 4 Refer to the Environmental Conditions section of the Agilent GC GC MS and ALS Site Preparation Guide for more detail Instr...

Page 4: ...de for more detail Oven type Heat dissipation Standard oven ramp 7681 BTU hour maximum Fast oven ramp options 002 and 003 10 071 BTU hour maximum 10BVenting the Oven Oven Heat Deflector Option 306 or Part Number G1530 80650 Below is a picture that shows the back view of an installed 7890 GC with the Oven Heat Deflector installed The exhaust duct is 10 cm 4 inches in diameter and adds 14 cm 5 5 inc...

Page 5: ...nt uncombusted hydrogen from the detector and from the inlet split and septum purge vents The uECD exhaust vents through a stainless steel tube connected to a length of large I D tubing that exits the back panel This should be routed to a fume hood or appropriate venting system Agilent Technologies recommends a vent line internal diameter of 6 mm 1 4 inch or greater With a line of this diameter th...

Page 6: ...edicated Fast 200 Japan Split Phase 48 63 2950 14 8 15 Amp Dedicated Notes 1 The number and type of electrical outlets depends on the size and complexity of your system For example a GC system with a computer monitor printer and HUB Switch requires 5 outlets 2 The outlet for the GC must be dedicated to the GC with a dedicated ground 3 The GC will have a label next to the power cord connector that ...

Page 7: ...2 5m China 220 Volts 15 Amps GB 1002 4 5m Europe Korea 220 230 240 10 Amps CEE 7 V11 2 5m Denmark Switzerland 230 Volts 16 Amps SWISS DENMARK 1302 2 5m India South Africa 240 Volts 15 Amps AS3112 4 5m Israel 230 Volts 16 Amps 16 AWG ISRAELI SI32 2 5m Japan 200 Volts 20 Amps NEMA L6 20P 4 5m United Kingdom Hong Kong Singapore Malaysia 240 Volts 13 Amps BS89 13 2 5m United States 120 Volts 20 Amps 1...

Page 8: ...ogen Nitrogen and Argon methane 5 Volatiles inlet VI is calibrated for only Helium and Hydrogen 5 For GC MS requirements refer to the GC MS Gas and Reagent Requirements section in the Agilent GC GC MS and ALS Site Preparation Guide Detector Carrier gas Make up 1st choice Make up 2nd choice Purge or reference Electron capture Hydrogen Helium Nitrogen Argon methane Argon methane 5 Argon methane 5 Ni...

Page 9: ...NPD TCD ECD FPD Hydrogen 35 100 35 100 45 100 Air 55 100 55 100 100 120 Make up 55 100 55 100 55 100 55 100 55 100 Reference 55 100 31BAuxiliary EPC and Pneumatic Control channels The minimum supply pressure for AUX and PCM modules is 20 psi greater than pressure used in your method For example if you need a pressure of 20 psi for the method the supply pressure must be at least 40 psi AUX EPC PCM ...

Page 10: ...fer to the GC Installation Kits and GC Plumbing sections of this checklist for Part Numbers 4 Never use liquid thread sealer to connect fittings Never use chlorinated solvents to clean tubing or fittings 5 Agilent also recommends using traps to remove water hydrocarbons and oxygen or a combination trap such as the Gas Clean Filter System that removes all three Special Notes 1 Shutoff Valves are re...

Page 11: ...7157 0210 1 8 inch Ball Shutoff Valve for Carrier Gas Supplies order 1 for each inlet system 0100 2144 PTFE tape Never use liquid thread sealer to connect fittings 0460 1266 Miscellaneous Gas Plumbing Information 1 Cryogenic cooling with Liquid N2 requires 1 4 inch insulated copper tubing 25 30 PSI supply 2 Cryogenic cooling with Liquid CO2 requires 1 8 inch heavy walled stainless steel tubing 750...

Page 12: ... instrument Agilent highly recommends the G3388B Leak Detector or equivalent to safely check for leaks Supply tubing for Hydrogen Gas Agilent recommends using NEW chromatographic quality copper or stainless steel tubing and fittings when using hydrogen Do not re use old tubing when installing or switching to hydrogen carrier gas Hydrogen gas tends to remove contaminants left on old tubing by previ...

Page 13: ...the GC system installation go smoother All kits include two Shutoff valves one for each inlet carrier supply Description Part number Installation Kit for FID NPD FPD Includes Gas Clean Filter Kit CP736538 for Moisture O2 and Hydrocarbon removal 19199N Installation Kit for TCD ECD MSD no Gas Filters Included order separately for ECD Gas Clean Filter is included with MSD 19199M ...

Page 14: ...ies Inlet septa appropriate for type Injection port seal Inlet insert or liner Injection port Capillary Column Ferrules Graphite Column installation 2 Capillary Column Nuts Column installation 2 Column Hangers Column installation Gas ID Labels For Labeling Gas Supply Tubing to Inlets and Detectors LAN Cable Communication to the GC Recommended Tools for GC Maintenance Tool Used for GC Tool Kit 5182...

Page 15: ...ning kits Application supplies Standards columns syringes Sampler supplies Vials caps electronic crimpers and syringes 2BAutosampler Hardware from Older Systems If you previously purchased samplers and would like to use these on your new GC the samplers may need firmware updates Sampler models that are compatible include 7683A and 7693A ALS and 7694B and 7697A Headspace Samplers This information i...

Reviews: