7890 Series GC Site Preparation Checklist
Issued: 09-Dec-2015, Revision: 1.8
Copyright
2013
Agilent Technologies
Page 2 of 15
18B
19B
Laboratory Bench Space - Dimensions and Weight
Identify the laboratory bench space before your system arrives based on the table below.
Pay special attention to the
total height and total weight requirements for all system components you
have ordered and avoid bench space with overhanging shelves
. Also pay special attention to the total
weight of the modules you have ordered to ensure your laboratory bench can support this weight.
7B
Special Notes
1.
Allow at least 25 cm clearance between back of GC and wall to dissipate heated air. See picture
below. A simple system that includes a GC and a computer requires about 86 cm of bench space.
2.
Avoid bench space with overhanging shelves. A 7683 or 7693 automatic liquid sampler will add to
the height of the instrument as shown below.
3.
G1888A Headspace, 5977 GCMS and QQQ MS are installed to the left of the 7890 and the 7697
Headspace and 220/240 Ion Trap MS are installed to the right. Refer to the "Dimensions and
Weight" section of the "Agilent GC, GC/MS, and ALS Site Preparation Guide" for more detail.
Instrument Dimensions
Component
Height (cm)
Width (cm)
Depth (cm)
Weight (kg)
G3440A Agilent 7890 GC
50 to 58
59
54
50
G3440A with 3rd detector
50 to 58
68
54
57
G2913A 7683 Auto-injector
42 above GC
3.1
G2614A 7683 Tray
30 Left of GC
3.0
G4513A 7693 Auto-injector
50 above GC
3.9
G4514A 7693 Tray
45 Left of GC
2 cm in front of
GC
6.8
7890 GC with 7683 ALS System
7890 GC with 7693 ALS System
Conversions: 1 kg = 2.2 pounds; 1 cm = 0.39 inches.