Operating the 708-DS / 709-DS
4
708-DS / 709-DS Operator’s Manual
91
From the Diagnostics screen:
•
Press
Next
to proceed to the second Diagnostics screen.
•
Press
Finish
to return to the Menu screen.
Manifold Control
The manifold may be ejected for cleaning and maintenance. Before
ejecting, raise the head, disconnect the two thermistor cables, remove
the tubing from the tubing clamps, and remove the evaporation covers.
Press
Eject
to lower the manifold until it is free. To reattach the
manifold, insert it into the guides and gently lift it up while pressing
Load
. Release the manifold once it begins to rise on its own. Press
Stop
to stop all manifold motion.
Dosage Delivery
Module Control
The DDMs are solenoid activated doors that hold a tablet above each
vessel. Press the corresponding button to open a DDM. Press
All
to
open all DDMs simultaneously.
Drive
Press
Drive
to raise the drive unit. Keep pressing it to raise the drive
unit to the maximum height. Release the button to stop the motion. Press
Drive
to lower the drive unit.
Manifold
Press
Manifold
to raise the manifold to the maximum height. Press
Manifold
to lower the manifold to the appropriate position based on
the apparatus type and volume.
Table 10
Menu Screen - Diagnostics Screen 1
Option
Function