80
708-DS / 709-DS Operator’s Manual
4
Operating the 708-DS / 709-DS
Menu Screen - Alarms
Two different alarms are available through the Alarms screen:
Table 5
Menu Screen - Alarms
Alarm
Function
Timer
This alarm functions as a countdown timer. Enter the duration as
HH:MM:SS. Press
Ok
to start the timer. When the alarm duration
expires, the Alarm Time Expired warning displays (
on page 106) and an audible alarm sounds. This alarm sounds
until you press
Ok
. To cancel a previously running alarm before time
has expired, enter 0 for the Timer Alarm value.
Pause Duration
This field allows you to set the maximum pause period while running a
method. When a running method is paused, a timer starts. If the pause
time exceeds this Pause Duration, the Pause Duration Exceeded error
displays (
on page 110), an alarm sounds,
and a message is sent to the printer. Entering a zero in this field disables
the Pause Duration Alarm.