124
708-DS / 709-DS Operator’s Manual
5
Maintenance and Troubleshooting
Water Bath Temperature Probe Accuracy Test
The temperature probe can be tested using a high-quality ohmmeter
capable of at least four-digit resolution and a known temperature bath.
Both the multi-meter and the bath temperature must be traceable to a
known reference standard such as NIST.
1
Place the probe into a water bath with a known temperature and
allow several minutes for the probe to equilibrate.
2
The resistance is measured by attaching the leads of the multi-meter
to the top and shank of the ¼-inch phone plug.
3
Refer to the table on Page
to find the resistance value of the
probe at the bath temperature. Temperatures between the values
listed may be interpolated.
4
Probes are interchangeable and manufactured with a tolerance of +/-
0.1 °C. Probes found to be outside of the tolerance of +/- 0.1 °C
should be replaced.