Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4
2-7
Driver Function Reference
Parameters
Table 2-3
VSU output voltage, resolution, and compliance
current
0 <
range
≤
10 pA (for HRSMU)
10 pA limited auto ranging
c
10 pA <
range
≤
100 pA (for HRSMU)
100 pA limited auto ranging
c
100 pA <
range
≤
1 nA
1 nA limited auto ranging
c
1 nA <
range
≤
10 nA
10 nA limited auto ranging
c
10 nA <
range
≤
100 nA
100 nA limited auto ranging
c
100 nA <
range
≤
1
μ
A
1
μ
A limited auto ranging
c
1
μ
A <
range
≤
10
μ
A
10
μ
A limited auto ranging
c
10
μ
A <
range
≤
100
μ
A
100
μ
A limited auto ranging
c
100
μ
A <
range
≤
1 mA
1 mA limited auto ranging
c
1 mA <
range
≤
10 mA
10 mA limited auto ranging
c
10 mA <
range
≤
100 mA
100 mA limited auto ranging
c
100 mA <
range
≤
1 A (for HPSMU)
1 A limited auto ranging
c
a. Negative
range
values are available for the functions used to start or execute measurements. The
negative values set the ranging mode to the fix, not the limited auto.
b. SMU uses the optimum range to force/measure voltage or current.
c. SMU uses the optimum range to force/measure voltage or current. Then, the SMU never uses
the range less than the specified range.
Output range
(actually used)
Output voltage
a
a. Parameter name may be
base
,
bias
,
peak
,
stress
,
value
,
start
,
stop
, and so on.
Setting resolution
comp
value
b
b. Set any value (e.g. 0.1) though the compliance is always 100 mA.
20 V
0
≤
|V|
≤
20 V
1 mV
±100 mA
Voltage or
current
Available
range
values
a
Ranging mode used for
output/measurement
Summary of Contents for 4155C
Page 3: ......
Page 13: ...1 Installation...
Page 19: ...2 Driver Function Reference...
Page 89: ...3 Programming Examples for Visual Basic Users...
Page 147: ...4 Programming Examples for Visual Basic NET Users...
Page 183: ...5 Programming Examples for C Users...
Page 225: ...6 Programming Examples for VEE Users...