Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4 3-31
Programming Examples for Visual Basic Users
Multi-Channel Pulsed Spot Measurements
Multi-Channel Pulsed Spot Measurements
This section explains example subprograms that enable/disable measurement
channels (perform_meas), perform multi channel pulsed spot measurement
(mspot_meas), and display measurement result data (display_data). This example
measures bipolar transistor collector current and base current.
Table 3-8
Multi-Channel Pulsed Spot Measurement Example
Sub perform_meas(vi As Long, ret As Long)
’1
Dim pins(3) As Long ’SMU port numbers
’3
pins(0) = 1
’SMU1: emitter
pins(1) = 2
’SMU2: base
pins(2) = 4
’SMU4: collector
ret = hp4156b_setSwitch(vi, pins(2), 1)
’8
ret = hp4156b_setSwitch(vi, pins(1), 1)
ret = hp4156b_setSwitch(vi, pins(0), 1)
check_err vi, ret
’11
mspot_meas vi, ret, pins()
’13
ret = hp4156b_setSwitch(vi, hp4156b_CH_ALL, 0)
’15
check_err vi, ret
End Sub
’18
Line
Description
1
Beginning of the perform_meas subprogram.
3 to 6
Declares variables, and defines the value.
8 to 10
Enables measurement channels.
13
Calls the mspot_meas subprogram (next page) to perform multi
channel pulsed spot measurement.
15
Disables measurement channels.
11 and 16
Calls the check_err subprogram (shown in Table 3-1) to check if
an error status is returned for the previous line.
18
End of the perform_meas subprogram.
Summary of Contents for 4155C
Page 3: ......
Page 13: ...1 Installation...
Page 19: ...2 Driver Function Reference...
Page 89: ...3 Programming Examples for Visual Basic Users...
Page 147: ...4 Programming Examples for Visual Basic NET Users...
Page 183: ...5 Programming Examples for C Users...
Page 225: ...6 Programming Examples for VEE Users...