Installing the NPD/ECD Chemical Filters Accessory 19199N

Preparing the Instrument

3

Note

The filter(s) can be installed on the flow panel in either a front or back position.

8.

If an auxiliary flow panel is not installed, loosen the screws in the front 
or back mounting location on the flow panel. Slide the chemical filter 
mounting bracket and filter assembly onto the screws, then tighten.

9.

If an auxiliary flow panel is present:

a.

Loosen the flow panel mounting screws.

b.

Slide the chemical filter mounting bracket (from the top) between 
the pressure regulators and the auxiliary flow panel. Position the 
top hole in the chemical filter bracket over the top mounting screw 
of the auxiliary flow panel.

c.

Tighten the auxiliary flow panel mounting screws.

Chemical filter assembly

Chemical filter
mounting bracket

Front mounting
location

Rear mounting
location

Summary of Contents for 19199N

Page 1: ...Installation Guide NPD ECD Chemical Filters Accessory 19199N...

Page 2: ...e unit against unintended operation Refer servicing to qualified service personnel Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard Discon...

Page 3: ...dangerous shock hazard by disconnecting the power cord before working on the instrument 1 Set the main power line switch to the off position 2 Disconnect the power cable from its receptacle 3 Allow ti...

Page 4: ...at hold it in position then sliding the tray away from the instrument b Lift the autosampler tray from its mounting bracket and set it aside c Remove the autosampler bracket by removing the 6 screws s...

Page 5: ...ing bracket and filter assembly onto the screws then tighten 9 If an auxiliary flow panel is present a Loosen the flow panel mounting screws b Slide the chemical filter mounting bracket from the top b...

Page 6: ...from the filters Route these lines to the detector flow manifold WARNING Hydrogen H2 is flammable and is an explosion hazard when confined in an enclosed space such as the oven In any application usi...

Page 7: ...from the side of the detector flow manifold Remove the flow lines from the detector flow manifold 12 Connect the lines from the detector chemical filters to the detector flow manifold using the outlet...

Page 8: ...sory 19199N Preparing the Instrument 6 13 Turn on the supply gases connected to the Detector Support Gas fittings Leak test the connections at the flow manifold block and the top of the detector chemi...

Page 9: ......

Page 10: ...Printed on recycled paper This product is recyclable Agilent Technologies Inc Printed in USA Dec 2000 19231 90767...

Reviews: