6
Oscilloscope Utility Settings
144
Agilent 1000B Series Oscilloscopes User’s Guide
Running Self-Calibration
The automatic calibration routine adjusts the internal circuitry of the
oscilloscope for the best measurement accuracy.
The automatic calibration should be run when the ambient temperature
changes by 5
°
C or more.
To run the oscilloscope’s self- calibration:
1
Press
[Utility]
.
2
In the Utilities menu, press
Self-Cal
.
3
Follow the instructions on the Calibration screen.
N O T E
Before performing the automatic calibration, let the oscilloscope warm-up at least 30
minutes.
Figure 58
Calibration Screen