3
Capturing Data
96
Agilent 1000B Series Oscilloscopes User’s Guide
To specify a trigger holdoff
Trigger holdoff can be used to stabilize a waveform. The holdoff time is
the oscilloscope's waiting period before starting a new trigger. The
oscilloscope will not trigger until the holdoff time has expired.
To specify a trigger holdoff:
1
Press
[Menu]
.
2
In the Trigger menu, press
Set Up
.
3
In the Set Up menu, press
Holdoff
and turn the
entry knob to adjust
the holdoff setting.
To reset the trigger holdoff
1
In the Set Up menu, select the
Holdoff Reset
menu item to return the
trigger holdoff setting to the 100 ns minimum value.
Figure 38
Trigger Holdoff