1
Getting Started
26
Agilent 1000B Series Oscilloscopes User’s Guide
Step 6. Compensate probes
Compensate probes to match your probe to the input channel. You should
compensate a probe whenever you attach it for the first time to any input
channel.
Low Frequency Compensation
For the supplied passive probes:
1
Set the Probe menu attenuation to 10X. If you use the probe hooktip,
ensure a proper connection by firmly inserting the tip onto the probe.
2
Attach the probe tip to the probe compensation connector and the
ground lead to the probe compensator ground connector.
3
Press the
[Auto Scale]
front panel key.
4
If waveform does not appear like the Correctly Compensated waveform
shown in
, then use a nonmetallic tool to adjust the low
frequency compensation adjustment on the probe for the flattest square
wave possible.
Figure 4
Low Frequency Probe Compensation
Low frequency compensation adjustment
Over Compensated
Under Compensated
Correctly Compensated