Step 18. Operator’s Check
After upgrading your instrument, complete the following test to verify its performance.
1. Turn the power on.
2. Perform the Operator’s Check. Refer to the Operating Reference 8757D Scalar Network
Analyzer (08757-90169) on page A-1. The Operator’s Reference is part of the Operating
Manual (08757-90109).
Step 19. Troubleshooting
If there is a possibility that the log amplifier boards (white guides) were not replaced in their
original positions, perform the following procedures.
1. Perform the Calibration Check Procedure. Refer to the Operating and Service Manual for
the 11613B Calibrator (11613-90011). Follow the procedure given for the 8757A in
Chapter 3.
2. If the Calibration Check Procedure fails, perform the Calibration Procedure. Refer to the
Operating and Service Manual for the 11613B Calibrator (11613-90011). Follow the
procedure given for the 8757A in Chapter 3.