31
2241A EN 20040210
3
Place the cassette you want to give emergency status first in the stack of
cassettes in the input buffer.
#
Do not place the cassette with emergency status in the stack while the
input mechanism is busy getting a cassette from the stack.
#
If you do not enter a cassette within 1 minute after pressing the
Emergency key, the CR 75.0 will quit the emergency function and return
to the operator main screen.
If fast preview is enabled, the CR 75.0 transmits the digital image data in
blocks of typical 100 lines to the preview station.
When the CR 75.0 has read the identification data of the emergency
cassette, it displays the operator main screen. The Digitizer resumes
processing the remaining cassettes in the cassette input buffer.
If you decide not to assign emergency status to a cassette after having
pressed Emergency, you can quit the Emergency function by either pressing
Escape or by pressing the Emergency key a second time (‘toggle’ key).
#
If a ‘WARNING’ or ‘LOCKED’ message is displayed during the
Emergency procedure, the CR 75.0 will not quit the Emergency mode.
Refer to
Chapter 4, ‘Preventive maintenance and troubleshooting’
.
4
Remove the cassette from the output buffer.