Remarks for HF-emission and immunity
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
RF Emission
Measurements
Agree-
ment
Electromagnetic Environment Guidelines
High frequency
RF emissions in
accordance with
CISPR 11
Group 1 The device uses high frequency energy exclu-
sively for its internal functions. For this reason,
its high frequency RF emission is very low and
it is improbable that neighboring electronic
equipment will be disrupted.
High frequency
RF emissions in
accordance with
CISPR 11
Class A
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required) this equipment
might not offer adequate protection to radio-
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Harmonic emis-
sion in accord-
ance with IEC
61000-3-2
Class A
Voltage fluctua-
tions / flickering
in accordance
with IEC
61000-3-3
Fulfilled
The device is intended for use in a professional healthcare facility
environment. Environmental conditions are stated in the user manual.
This device was tested for a professional healthcare environment as described
above. Nevertheless the HF-emission and immunity can be influenced by
connected data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Resistance to
Jamming Test
IEC 60601-1-2:2014
Test Level
Level of
Agreement
Electromagnet-
ic Environment
Guidelines
Discharge of
static electricity
in accordance
+ 8 kV contact dis-
charge
+ 8 kV contact
discharge
Floors should
consist of wood,
concrete or ce-
Full Leg Full Spine DR Retrofit System | Technical Data |