Execute the exposures with external collimator
In the operator room:
1.
Press the exposure button to acquire the first partial image.
The partial image is sent to the NX workstation.
2.
Press the button on the Full Leg Full Spine control unit.
The DR detector is moved to the next position inside the DR detector
holder. The external collimator is also moved to the next position.
3.
Repeat previous steps to move the DR detector to each required position
and acquire the partial images.
If only two partial images are required and the examination contained
three partial image thumbnails, the FLFS workflow can be cancelled after
the second exposure. The stitching must be started manually.