Starting the DR Full Leg Full Spine application
Note
:
Connecting the external collimator while the Full Leg Full
Spine wall stand is powered on, will lead to a restart of the Full
Leg Full Spine wall stand together with the external collimator.
First connect the external collimator to the Full Leg Full Spine
wall stand; then connect the wall stand to the power supply.
1.
Connect the external collimator to the Full Leg Full Spine wall stand.
2.
Connect the Full Leg Full Spine wall stand to the power supply.
WARNING:
Because the equipment cable is long, take care that cables do
not become tangled during use. Also, be careful not to get your
feet caught in the cable. It may cause a malfunction of the
equipment or injury to the user from tripping over the cable.
38
| Full Leg Full Spine DR Retrofit System | Basic Workflow
0326A EN 20211014 1141