WARNING:
When operating or maintaining the printer, always observe
the following safety guidelines:
• Have electrical or mechanical defects repaired by skilled
personnel only!
• Do not modify this equipment without authorization of the
manufacturer.
• Do not override or disconnect the integrated safety features.
• Ventilation openings may not be covered.
• Position the printer in a way that it could be easily
unplugged in order to achieve separation from mains.
• Always switch off the printer and disconnect the power cord
from the outlet before carrying out any maintenance work.
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
The classification of this product according to the medical electrical
equipment standard IEC 60601-1 requires installation outside the patient
vicinity. For definition of patient vicinity see dimensions below.
1.
R = 1.5 m/4.9 feet (EN 60601-1) or 1.83 m/6 feet (UL 60601-1).
2.
h = 2.5 m/8.2 feet (EN 60601-1) or 2.29 m/7.5 feet (UL 60601-1).