WARNING:
Make sure to measure distance A from the left edge of
the left line to the right edge of the right line and
distance B from the upper edge of the upper line to the
lower edge of the lower line.
We strongly recommend using a 30 cm (12-inch) machinist
scale with 0.5 mm divisions (1/64 inch).
3.
Record these values as measured distances A and B on Chart 4 (‘Geometric
Consistency Control Chart’).
4.
Compare the measured A and B dimensions with the reference dimension
values, A
ref
and B
ref
on Chart 4 (‘Geometric Consistency Control Chart’).
The differences between the measured dimensions of A and B and the
reference values A ref and B ref should be less than or equal to 1.0%.
5.
Check for image distortion.
6.
Calculate the aspect ratio by dividing A by B.
The result must be 1 +/- 0.01
WARNING:
If the image size or distortion values are outside of limits,
contact your local service organization to resolve the
problem.
Related Links
Establishing the image geometry reference values for mammography application
on page 103
on page 135
DRY 3.4 | Advanced Operation (key-operator mode) |
109
2602A EN 20191018 0826