Establishing the daily operating reference density levels for
mammography application
WARNING:
The densitometer of the printer is calibrated at installation.
Authorized service personnel should recalibrate the internal
densitometer annually or after major service or repair.
This procedure enables you to establish the base line values for:
• Base + Fog density
• Low density
• Mid density
• High density
To establish the daily operating levels, proceed as follows:
1.
Press the Key-operator key to enter the Key-operator mode.
The ‘Key-operator main menu’ screen appears:
2.
Press the Down key eight times, followed by the Confirm key to select
‘Quality Control’.
The ‘Select input tray’ screen appears:
Note
:
When controlling the printer via a remote PC, The ‘Select
input tray’ screen is preceded by a screen, which allows you
to:
• Start the quality control procedure immediately,
• Edit additional data for the last quality control
measuring.
3.
Press the Up/Down arrow keys to select the proper input tray, followed by
the Confirm key.
The printer will automatically print the QC mammography test image.
4.
After the image is printed, the system will display all measured optical
density values:
DRY 3.4 | Advanced Operation (key-operator mode) |
101
2602A EN 20191018 0826
Summary of Contents for 5364/500
Page 1: ...DRY 3 4 5364 500 User Manual 2602A EN 20191018 0826 ...
Page 12: ...12 DRY 3 4 DRY 3 4 2602A EN 20191018 0826 ...
Page 58: ... Audio signals The keypad The display 58 DRY 3 4 Introduction 2602A EN 20191018 0826 ...
Page 87: ...The menu structure DRY 3 4 Advanced Operation key operator mode 87 2602A EN 20191018 0826 ...