Chapter 4 Maintenance Procedures
114
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
4.
Insert the open end of the waste tubing (the end without the quick-disconnect fitting)
into a sink or container that is lower than the level of the Vac Waste port (the waste
water drains by gravity).
WARNING
!
WARNING!
If you are using an external waste tank, the waste tubing is already plugged into the
Waste port. Pull it out of the Waste port and connect it to the Vac Waste port. After the vacuum
tank is drained, connect the waste tubing back to the Waste port.
5.
Insert the waste tubing quick-disconnect fitting into the Vac Waste port. Push the
fitting in until it clicks into place
.
CAUTION
!
Waste water will flow immediately when the waste tubing is connected to the Waste Vac port. Be
prepared for this. Only connect the quick-disconnect fitting to the Waste Vac port. Do not insert
the open end of the waste tubing into the port.
WARNING
!
WARNING!
The waste water contains ethanol. Dispose of it properly as a flammable liquid. Refer
to the MSDS provided by your ethanol supplier for disposal information.
Figure 4.37 Open the Vacuum Vent Switch
Figure 4.38 Waste Tubing
Vacuum vent
switch
Open end
Quick-disconnect fitting