Chapter 4 Maintenance Procedures
110
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
CAUTION
!
Make sure the filter end of the tubing does not come into contact with anything that might
contaminate it.
16. Drain the water from the supply tank:
a. Tap
SUPPLY TANK Drain
on the Maintenance screen.
An on-screen message appears (
).
b. Tap
OK
in the on-screen message.
An on-screen message appears during the draining process (
).
The water in the supply tank is flushed through the wash system and empties into
the waste tank. Wait for the supply tank to drain.
17. Fill the supply tank again with Type 1, >18.2 M
Ω
water following the procedure on
. After the supply tank is filled, remove the supply tubing from the supply port.
Figure 4.31 Draining the Supply Tank
Figure 4.32 On-screen Message – Supply Tank Draining