FAILURE MODE
Determines the current value at the output of the product in the event of an echo loss ("NO
ECHO").
MEDIUM TEMPERATRE
FAILURE MODE
HART
FACTORY DEFAULT
RESET
FAILURE MODE
NO ECHO: 3.75 mA
DEAD ZONE: 22 mA
NO ECHO: Current in case of echo loss
DEAD ZONE: Dead zone current
The values can be set in three steps:
3.75 mA, 22 mA and LAST (last mea-
sured data).
HART
This menu item belongs to the menu of the product with the current output UST-20. HART®-Pro
-
tocol setting (point to point, multidrop) and address for the multidrop mode. Up to 15 units can be
connected to one two-wired cable in the multidrop mode.
MEDIUM TEMPERATRE
FAILURE MODE
HART
FACTORY DEFAULT
RESET
HART:
POLLING ADDRESS
00
In case of the address "00", the point to
point mode is enabled. The range from
"01" to "15" is reserved for addresses in
the multidrop mode.
MODBUS
This menu item belongs to the menu of the product with the output Modbus UST 20. Modbus
mode is intended for the settings of the level Modbus addresses, baud rate and parity settings.
MEDIUM TEMPERATRE
FAILURE MODE
HART
FACTORY DEFAULT
RESET
MODBUS:
ADDRESS: 1 to 247 (default 1) 4800, 9600,
19200 (default 9600)
FORMAT : 8N1, 8O1, 8E1, 8N2 (default 8N1)
FORMAT: number of Stopbits:
FACTORY DEFAULT
To reset the initial values of the level meter set by the manufacturer, press the OK button (see the
Factory default table).
MEDIUM TEMPERATRE
FAILURE MODE
HART
FACTORY DEFAULT
RESET
FACTORY DEFAULT
ARE YOU SURE?
After you press the OK button, "RUN
-
NING" will be displayed for about 3 sec.
After the initial values are set, "DONE"
will be appear on the display.
FACTORY DEFAULT
DONE
PRESS TO EXIT
RUNNING
FACTORY DEFAULT
ADDRESS: 001
BAUD RATE: 9600
FORMAT: 8N1
parity
N – non parity
O – odd parity
E – even parity
data: 8 - number of bits
Summary of Contents for SonarFox UST 20
Page 2: ...SonarFox UST 20 AFRISO 2 ...
Page 34: ...SonarFox UST 20 AFRISO 34 20 EU Declaration of Conformity ...
Page 35: ...35 AFRISO SonarFox UST 20 ...
Page 36: ...SonarFox UST 20 AFRISO 36 21 EU Type Examination Certificate ...
Page 37: ...37 AFRISO SonarFox UST 20 ...
Page 38: ...SonarFox UST 20 AFRISO 38 ...