Installation in standpipe
Fig. 13: Installation in standpipe
By using a standpipe, the effects of tank instal
-
lations, foam formation and turbulence can be
prevented.
The standpipe must be made of a material
with a smooth inner surface. After installation,
you must perform a false signal suppression
using the "LEARN" menu item.
When using pipe sockets that protrude into the
closed container, the stand pipe must end at a
45° angle (Fig. 13 righthand).
Measurement in the standpipe is not suitable
for bulk solids that tend to adhere strongly.
ventilation holes