SonarFox UST 20
AFRISO
16
8 . Settings
The product is manually controlled using 3 buttons located on the
removable display module.
The text "SAVED" at the bottom of the display indicates that the values
have been saved. Values not confirmed by pressing button will not be
saved! After 5 minutes of inactivity, the level meter automatically switches
back to the measuring mode. If the password is activated, the level meter
will also lock itself. Once it is locked, it is not possible to make any
changes to the settings. When any editing is attempted, the display will
show the text "NO PASSWORD". The unlocking procedure is described
on page 23.
After connecting the supply voltage, the product display will show the manufacturer's logo and
the text "Starting" (approx. 30 s). Then the menu switches to the measuring mode and the display
shows the current measured value.
Basic settings
The basic setting is required after the product has been commissioned
for the first time (Setting the measuring range, selecting the units or
damping, sensitivity and teaching).
MIN LEVEL and MAX LEVEL
The minimum / maximum distance to the level is set here („LEVEL“ for currents 4 mA / 20 mA),
Then the value can be assigned on the display under the menu item "DISPLAY". The units are set
in the "UNITS" menu.
MIN
CURRENT LEVEL
02870mm
OUTPUT:
04.00
mA
LEVEL
06000
mm
DISPLAY:
00000
%
MAX
CURRENT LEVEL
02870mm
OUTPUT:
20:00 mA
LEVEL
00250 mm
DISPLAY:
00100 %
MIN LEVEL
MAX LEVEL
UNITS
DAMPING
SENSITIVITY
TEACHING
MIN LEVEL
MAX LEVEL
UNITS
DAMPING
SENSITIVITY
TEACHING
• CURRENT
: currently measured
distance to the level
• OUTPUT
: Current 4 mA / 20 mA
• LEVEL
: Definition OF MIN/MAX
distance to the level
• DISPLAY
: Value on the display
If the message "OUT OF LIMITS" appears in the lower part of the display when entering values,
then the value entered in the "LEVEL" menu item is outside the measuring range. If the message
"PAN TOO SMALL" appears, enter a larger range of min / max values, see chapter. "Specifica
-
tions".
The decimal point position is fixed in the menu item "LEVEL" is fixed in the menu item "LEVEL"
(based on the selected units). The position of the menu item “DISPLAY” is freely adjustable.
1.
Press the OK button to open the "BASIC SETTINGS" menu.
2.
Use the arrow buttons to select the "MIN LEVEL" or "MAX LEVEL" menu and press OK.
3.
Press the OK button and the arrow buttons to set the distance for the defined current "LEVEL".
The value is shown on the "DISPLAY" display.
4. Press the OK button to save the values.
5.
Press the ESC button to exit the menu. The product switches back to the measurement mode.
STARTING
AFRISO
®
UNITS
UNITS
mm
DISPLAY:
%
SAVED
BASIC SETTINGS
SERVICE
DIAGNOSTIC
CLONE SETTINGS
PASSWORD
LANGUAGE
INFO
Summary of Contents for SonarFox UST 20
Page 2: ...SonarFox UST 20 AFRISO 2 ...
Page 34: ...SonarFox UST 20 AFRISO 34 20 EU Declaration of Conformity ...
Page 35: ...35 AFRISO SonarFox UST 20 ...
Page 36: ...SonarFox UST 20 AFRISO 36 21 EU Type Examination Certificate ...
Page 37: ...37 AFRISO SonarFox UST 20 ...
Page 38: ...SonarFox UST 20 AFRISO 38 ...