Table 3-14:
Limit Switch Specifications
Specification
Supply Voltage
5 V ±5%
Supply Current
25 mA
Output Type
Open Collector
Output Voltage
5 V
Output Current
10 mA (sinking)
Output Polarity
(Factory Configured)
Normally Closed (NC)
l
Sinks current to ground (Logic "0") when not in limit
l
High impedance (Logic "1") when in limit
l
Requires external pull-up to +5 V (10 kΩ recommended)
Note:
If the APR100DR is driven beyond the electrical limit, it will encounter a mechanical stop. Impacting the mechanical
stop could cause damage to the stage even at low speeds.
Table 3-15:
Brake Specifications
Specification
Supply Voltage
24 VDC
Supply Current
(1)
690 mA Typical
(1) Required to release the brake and allow motion.
Hardware Manual
APR100DR
www.aerotech.com
35