MRK-D-0034 V2.0
Aeroqual AQM 65 User Guide
Page | 44
3.4.2 Service and Calibration app
The Service and Calibration app has three functions, Service mode initiation, changing gain and offset
during calibration, and configuring and running the AirCal 8000.
The Service and Calibration app can be accessed by all user types and all instrument licences.
Service Mode button
When a service engineer visits the AQM 65 to perform routine service such as filer changes or flow
checks, this can cause the data to read very high or very low. During service activities the AQM 65
ca
n be put in to Service Mode by clicking the service mode button. This will tag the data as “service
data” which can later be filtered when daily or hourly averages are being calculated.
See Section 4.3.1for further discussion on Service Mode
Gain and offset adjustment
Gain and offset adjustments are made as part of the AQM 65 gas calibration procedure. These
adjustments are made in the Service and Calibration App.
AirCal 8000
The AirCal8000 is an optional integrated calibration module. The module can be programmed to
deliver gas from attached calibration cylinders. Gas can be delivered manually or a schedule can be
set to deliver calibration gas automatically.
Figure 3-14 The service and calibration app can be used to put the AQM in to service mode, adjust offset
and gain settings and control the AirCal 8000