MRK-D-0034 V2.0
Aeroqual AQM 65 User Guide
Page | 10
1.1.2 Gas flow through modules
The gas flow through all of the gas modules is powered by a single diaphragm pump at the bottom left
of the enclosure. The vacuum side of the pump is connected to the gas outlet (exhaust) of each
module. This is illustrated for a single gas module in Figure 1-5.
The pump pulls sample air through the module and then pushes air out through the exhaust so that
the sample air passes through the pump only
after
it has been measured in the module.
There are two different types of tubing material used to handle the gas flow.
Before the gas reaches the sensor, all of the tubing material is rigid non-reactive PFA.
After the sample has passed the sensor all of the tubing is flexible R3603 Tygon
Figure 1-5 A single pump is used for all gas modules. The vacuum port of the pump is connected to the
outlet of each gas module
The gas modules are connected in two rows inside the AQM 65 enclosure. There is one main gas
inlet for sampling ambient air and one gas exhaust port both shown in Figure 1-6. Each gas module is
connected to the pump via the
Tygon exhaust manifold
. The modules are connected in
parallel
as
shown in Figure 1-6 on the left side. Each gas module is connected to the inlet via the
PFA sample
manifold
, also in parallel as shown in Figure 1-6 on the right.