14. Printing Intervals Set

a) Enter “Printing Intervals”.

b) Click the existing time figure in “Printing Intervals” bar.

c) Input the suitable printing intervals by clicking the newly 

opened  setting box, in the range of 0 -60 minutes. Setting it at 

0 min will shut down this printing function.

d) On the printing slip, four key parameters can be displayed, 

Temperature, Humidity, Oxygen Density and Carbon Dioxide 

Density.

e) “Hospital or Clinic” names can be printed on the slip as a 

customization option; please contact your local supplier for 

more details.

f) Switch on printing function when needed by shifting OFF 

button to ON as displayed below.

15. Running Status

On the Running Status page, all real-time running 

status will be displayed.

16. CO

2

 Remover Replacement Reminder

Depending on varied types of soda lime powder 

from different suppliers, the effective validation 

time may be different; the operator can set 

17. Parameter Calibrations

The Parameter Calibration needs to be managed 

by trained qualified technician, or under the 

guidance and supervision of our technicians.

On this page, you can make adjustments to these 

parameters below.

Temperature Setting Zero and Full.

Humidity Setting Zero and Full.

Ambient Temperature and Auxiliary Temperature.

CO

2

 Density Setting Zero and Full.

25.0

OFF
OFF
OFF
OFF
OFF

30.0

50

1500

3

Auto

0
0
1
9
8

+

-

OFF

0

Prin�ng Intervals

1

2

3

4

5

6

7

8

9

+/-

0

.

CLR

DEL

Enter

0

60

30

25.0

OFF
OFF
OFF
OFF
OFF

30.0

50

1500

3

Auto

0
0
1
9
8

+

-

ON

30

CO

2

 Remover 

Replacement Reminder

RENEW

different replacement reminding times accordingly. 

At the bottom of the above page for “Running Status”, the estimated CO

2

 remover replacement time 

is displayed.

Click the boxes shown above to set effective time for soda lime CO

2

 remover reminder.

In box 4, it shows the elapsed working time from last replacement; box 2 is for resetting the validation 

time for different suppliers’ soda lime.

If you are using the soda lime from our company, please click “RENEW” button to set it at 800 hours 

after replacing.

Note: It’s highly recommended to use our soda lime to ensure a reliable working efficiency and 

service life.

Prin�ng Intervals

Prin�ng Intervals

18. History

 

The running status of four key parameters, 

temperature, humidity, CO

2

 density and O

2

 density, 

will be recorded every minute.

In our system, the operation history can be tracked 

for up to 72 hours /3 days as maximum.

All the curves will be displayed on this interface, you 

can click on any of them to enlarge the whole curve 

and get more detailed statistics.

19. Operation Guide

All the instructions shown above are also 

installed in the control panel system; they can 

be found in the Operation Guide page shown.

The instructions are displayed with detailed 

wording and photos.

20. Accessories Mount Installation

This stainless steel wire grid is designed for 

mounting all necessary accessories, such as IV 

pole, infusion pump, syringe pump, monitor, etc.

Please insert the bottom foot into the fixture hole 

on top of the large compartment first, then match 

and insert the top pin into the triangle connector 

between the two medium compartments as 

indicated by the picture.

21. Door Shade Curtain Installation

A slide track and four hooks are installed on top of 

the transparent doors on each compartment, to 

hang the shade curtains helping cut out light and 

noise transmissions to the compartment and calm 

the patients.

No other function can be set when the unit is being disinfected. The disinfecting process can be 

cancelled at any time returning the unit to its previous settings. All other lights will be cancelled and all 

functions will be stopped during the disinfection process.

O

2

 Density Setting Zero and Full.

Note: You need a pass code to enter this page, please contact your local supplier or service center 

for further assistance.

Summary of Contents for UC-1901BK

Page 1: ...Distributor Contacts UC 1801 KO2 10F KO2 2F UC 1901 Professional Multifunctional Veterinary Intensive Care Unit WMS 1501 Professional Vet Equipments Supplier User Manual...

Page 2: ...dimensions General usage guidelines General opreation tips Radio requency interferece Choosing the working space for the unit Parameters Operation guidance Setting interface Setting mode Trouble shoo...

Page 3: ...ge caused by a third party Transport or Delivery Company The day the unit is delivered an inspection of the outer packing should be made for any damage to the packing box Then the unit should be unpac...

Page 4: ...dB A Control Panel Screen Type Max Power Consumption Average Power Consumption Power Supply Failure Protection Negative Ion Generator Accessories Built in Printer Ultra violet Sterilization Blue Light...

Page 5: ...and displays 1 Setting Mode Allows the operator to switch on off and set all the changeable running parameters for most functions 2 Operating History Allowing the operator to view a complete operatio...

Page 6: ...rbon dioxide figure in CO2 Density bar c Input the desired carbon dioxide density by clicking the newly displayed setting box within the range of 200 2000ppm d Click Enter to set the new carbon dioxid...

Page 7: ...ight Therapy Exam Light Emergency Vent Se ng Blue Light Therapy MIDDLE 12 Nebulizer and Timer a Open the nebulizer chamber door on front control panel of the machine b Gently take the nebulizer assemb...

Page 8: ...er reminder In box 4 it shows the elapsed working time from last replacement box 2 is for resetting the validation time for different suppliers soda lime If you are using the soda lime from our compan...

Page 9: ...and right compartments will start to operate 1 Temperature Sensor Humidity Sensor Replacement Both Temperature and humidity sensors can be found can be found at the innermost top corner of the compart...

Page 10: ...he top cover on ICU compartment to expose internal components Please find the negative ion generator check if it s broken Open and remove the protective cover on the emergency ventilation fan on the b...

Reviews: