14. Printing Intervals Set
a) Enter “Printing Intervals”.
b) Click the existing time figure in “Printing Intervals” bar.
c) Input the suitable printing intervals by clicking the newly
opened setting box, in the range of 0 -60 minutes. Setting it at
0 min will shut down this printing function.
d) On the printing slip, four key parameters can be displayed,
Temperature, Humidity, Oxygen Density and Carbon Dioxide
Density.
e) “Hospital or Clinic” names can be printed on the slip as a
customization option; please contact your local supplier for
more details.
f) Switch on printing function when needed by shifting OFF
button to ON as displayed below.
15. Running Status
On the Running Status page, all real-time running
status will be displayed.
16. CO
2
Remover Replacement Reminder
Depending on varied types of soda lime powder
from different suppliers, the effective validation
time may be different; the operator can set
17. Parameter Calibrations
The Parameter Calibration needs to be managed
by trained qualified technician, or under the
guidance and supervision of our technicians.
On this page, you can make adjustments to these
parameters below.
Temperature Setting Zero and Full.
Humidity Setting Zero and Full.
Ambient Temperature and Auxiliary Temperature.
CO
2
Density Setting Zero and Full.
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
1
9
8
+
-
OFF
0
Prin�ng Intervals
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
60
30
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
1
9
8
+
-
ON
30
CO
2
Remover
Replacement Reminder
RENEW
②
③
④
different replacement reminding times accordingly.
At the bottom of the above page for “Running Status”, the estimated CO
2
remover replacement time
is displayed.
Click the boxes shown above to set effective time for soda lime CO
2
remover reminder.
In box 4, it shows the elapsed working time from last replacement; box 2 is for resetting the validation
time for different suppliers’ soda lime.
If you are using the soda lime from our company, please click “RENEW” button to set it at 800 hours
after replacing.
Note: It’s highly recommended to use our soda lime to ensure a reliable working efficiency and
service life.
Prin�ng Intervals
Prin�ng Intervals
18. History
The running status of four key parameters,
temperature, humidity, CO
2
density and O
2
density,
will be recorded every minute.
In our system, the operation history can be tracked
for up to 72 hours /3 days as maximum.
All the curves will be displayed on this interface, you
can click on any of them to enlarge the whole curve
and get more detailed statistics.
19. Operation Guide
All the instructions shown above are also
installed in the control panel system; they can
be found in the Operation Guide page shown.
The instructions are displayed with detailed
wording and photos.
20. Accessories Mount Installation
This stainless steel wire grid is designed for
mounting all necessary accessories, such as IV
pole, infusion pump, syringe pump, monitor, etc.
Please insert the bottom foot into the fixture hole
on top of the large compartment first, then match
and insert the top pin into the triangle connector
between the two medium compartments as
indicated by the picture.
21. Door Shade Curtain Installation
A slide track and four hooks are installed on top of
the transparent doors on each compartment, to
hang the shade curtains helping cut out light and
noise transmissions to the compartment and calm
the patients.
No other function can be set when the unit is being disinfected. The disinfecting process can be
cancelled at any time returning the unit to its previous settings. All other lights will be cancelled and all
functions will be stopped during the disinfection process.
O
2
Density Setting Zero and Full.
Note: You need a pass code to enter this page, please contact your local supplier or service center
for further assistance.