9. Exam Light
a) Enter “Setting Mode”.
b) Click the existing time figure in “Exam Light” bar.
c) Input the desired working time for the pure white exam light
by clicking the newly opened displayed setting box, in the range
of 1 -360 minutes.
d) Click “Enter” to set the exam light working time, click the “X” in
the top right corner of the settings box to close the settings box.
e) Switch on exam light when needed by shifting OFF button to
ON as displayed. The light can be switched off by shifting from
ON to OFF at anytime.
10. Blue Light Therapy
a) Enter “Setting Mode”.
b) Click the existing time figure in “Blue Light Therapy” bar.
c) Input the desired working time for blue light by clicking the
displayed setting box, in the range of 1 -360 minutes.
d) Click “Enter” to set the blue light working time, click the “X” in
the top right corner of the settings box to close the settings box.
e) Switch on blue light when needed by shifting OFF button to
ON as displayed below.
Remarks:
When therapeutic blue light is on, exam light and illumination
yellow light will be shut down automatically; once the blue light
finishes working, other lights will resume.
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
1
9
8
+
-
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
33
1
9
8
+
-
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
1
9
8
+
-
Exam Light
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
360
30
Blue Light Therapy
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
360
33
OFF
OFF
OFF
OFF
50
1500
3
Auto
中
0
1
9
8
+
-
8. Emergency Vent Setting
The emergency ventilation port is designed as a fail-safe, for
necessary fresh air in critical situations like a power supply
failure.
The fan is powered by a Lithium battery pack, which will allow several operations without a power
supply from the mains power.
a) Enter “Setting Mode”.
b) Select suitable vent opening mode by shifting the “ON”, “OFF” and “AUTO” on “Emergency Vent
Setting” bar.
Note: There are three working modes of Emergency Vent, ON, OFF and AUTO.
A. ON, means the vent will stay open unconditionally.
B. OFF, means the vent will be shut unless power supply failure is detected, then it will
be opened for fresh air.
OFF mode is NOT recommended to be used.
C. AUTO, means the vent will be opened in the following conditions.
C1: Power supply failure.
C2: O
2
Density in compartment drops to lower than 15%.
C3: O
2
Density ≤25% and CO
2
Density stabilizes at the level of greater than CO
2
Density Alarm setting for over 3 minutes.
C4: O
2
Density ≥25% and CO
2
Density level exceeds 4500ppm.
Blue Light
Therapy
Exam Light
Emergency Vent Se�ng
Blue Light
Therapy
MIDDLE
12. Nebulizer and Timer
a) Open the nebulizer chamber door on front control panel of
the machine.
b) Gently take the nebulizer assembly out from the chamber, be
careful not to stretch the pipeline too much.
c) Open nebulizer nozzle cap by twisting counterclockwise.
d) Pour medicine liquid into the medicine cup, replace nozzle
cap clockwise.
e) Attach the cup nozzle part into the silicon port at the left
upper corner of the transparent door.
f) Enter “Setting Mode”.
g) Click the existing time figure in “Nebulizer Timer” bar.
h) Input the desired working time for nebulizer by clicking the
newly opened displayed setting box, in the range of 1 -90
minutes.
i) Click “Enter” to set the nebulization working time; click the “X”
in the top right corner of the settings box to close the settings
box.
j) Turn the nebulizer on when needed by shifting OFF button to
ON as displayed below.
Remarks: During nebulization treatment, cooling, heating and
ventilation functions will shut down, for better preservation of
medicine atmosphere in the compartment.
13. Ultra-violet Timer
a) Enter “Setting Mode”.
b) Click the existing time figure in “Ultra-violet Timer” bar.
c) Input the desired working time for ultra-violet sterilization by
clicking the displayed setting box, in the range of 1 -90 minutes.
d) Click “Enter” to set the ultra-violet sterilization working time;
click the “X” in the top right corner of the settings box to close
the settings box.
e) Switch on ultra-violet sterilization light when needed by
shifting OFF button to ON as displayed below.
Note: There must be No Animals in the unit when this function is
being used.
11. Negative Ion Set
a) Enter “Setting Mode”.
b) Click the existing time figure in “Negative Ion Timer” bar.
c) Input the desired working time for negative ion gener-
ator by clicking the newly opened displayed setting box, in
the range of 0 -72 hours.
d) Click “Enter” to set the negative ion working time; click the
“X” in the top right corner of the settings box to close the
settings box.
e) Switch on negative ion generator when needed by shifting
OFF button to ON as displayed below.
Nozzle cap
Medicine cup
Cone gasket
Nebulizer Timer
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
90
20
Nega�ve Ion Timer
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
72
2
25.0
NO
NO
ON
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
2
9
8
+
-
25.0
NO
NO
ON
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
2
9
8
+
-
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
0
9
8
+
-
Ultra-violet Timer
1
2
3
4
5
6
7
8
9
+/-
0
.
CLR
DEL
Enter
0
~
90
15
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
1
9
0
+
-
25.0
NO
NO
NO
OFF
ON
30.0
50
1500
3
Auto
中
0
0
2
9
15
+
-
Nega�ve Ion
Timer
Ultra-violet Timer
Ultra-violet Timer
Nega�ve Ion
Timer
25.0
OFF
OFF
OFF
OFF
OFF
30.0
50
1500
3
Auto
中
0
0
0
9
8
+
-
Nebulizer Timer
Nebulizer Timer