9 - DC and AC CURRENT
(C.A 5220 only)
Always switch Off the circuit to be tested before
connecting the multimeter to the circuit
Warning:
connect the leads to the multimeter and connect in
series to the circuit
- the red lead to the + terminal up to 400mA
- the red lead to the 10A terminal, from 400mA to 10A
Place the selector on the A or mA ammeter function
Press the AC/DC button to select AC or DC
Energize the circuit to be tested
Automatic range selection : read the measured value
If necessary, press the following buttons
- RANGE to manually select the ranges
- MIN-MAX to read the min or max values
-
to light the display
DC currents
A DC
40mA
400mA
10A
(1)
Voltage
600mV
4.5V
600mV
drop
(2)
Digital
10µA
100µA
10mA
resolution
Bargraph
1mA
10mA
1A
resolution
Accuracy
± 1.5% of Reading ±1ct
± 1.5% of
Reading ±2cts
Protection
(3)
Fuse 0.4A HBC
Fuse 12A HBC
AC currents
A AC
40mA
400mA
10A
(1)
Voltage
600mV
4.5V
600mV
drop
(2)
Digital
10µA
100µA
10mA
resolution
Bargraph
1mA
10mA
1A
resolution
Accuracy
(4)
± 1.5% of Reading ±1ct
± 2% Reading
± 2cts
Protection
(3)
Fuse 0.4A HBC
Fuse 12A HBC
(1) 4000 count display limited to 10A Between 5A and 10A, to avoid overheating,
limit the operating time to 10 minutes
(2) Voltage drop at the terminals for 40mA, 400mA, 10 A
(3) Permitted overloads :
600mA for 30 seconds for the 400mA range
15A for 30 seconds for the 10A range
(4) Use on frequency : 40Hz to 500Hz