12
13-2 Storage
If the multimeter is not to be used for a period of 60 days or more,
remove the battery and store separately.
13-3 Cleaning
The multimeter must be disconnected from any
electrical source.
To clean the case, use a cloth slightly moistened with soapy water.
Rinse with a damp cloth. Then, dry rapidly with a cloth or in a hot air
stream.
13-4 Repair and Calibration
To guarantee that your instrument complies with the factory
specifications, we recommend that the Models C.A 5210 / C.A 5220
be submitted to our factory service center at one-year intervals for
recalibration, or as required by other standards.
For instrument repair and/or calibration, please call our factory, toll-
free, at
(800) 945-AEMC
(800-945-2362) :
CHAUVIN ARNOUX, Inc.
d.b.a. AEMC
Instruments
15 Faraday Drive
Dover, NH 03820 USA
Tel : (800) 954-2362
(603) 749-6434
Fax : (603) 742-2346
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to
N.I.S.T. are available upon request. Overseas customers must receive
written authorization before returning any instrument.
13-5
Technical and Sales Assistance
If you are experiencing any technical problems, or require any
assistance with the proper use or application of this instrument,
please call our technical hotline :
CHAUVIN ARNOUX, Inc.
d.b.a. AEMC
Instruments
200 Foxborough Blvd.
Foxborough, MA 02035
Tel : (800) 343-1391
(508) 698-2115
Fax : (508) 698-2118