When the guard terminal is
connected at “B”, the
instrument measures the
current “I” and not the
surface leakage current
“i
1
”, which is not included in
the measurement. See
Figure 9. This type of
measurement will give the
true value of the resistance
“Rx”, providing the “Ry”
and “Rz” are not too low.
Figure 9
Voltage Measurements (Safety Check)
These instruments can measure AC or DC voltages. For voltage
measurements, proceed as follows:
•
The push-to-test button should be in the OFF position.
•
Connect the instrument with the voltage leads connected between the
Line (–) and Earth (+) terminals on the unit.
•
Read the voltage directly on the voltage scale.
+ EARTH
— LINE
MEGOHMMETER
Figure 10
A
C
B
Ry
Rz
Rx
i1
i
To EARTH
Terminal
To LINE
Terminal
To Guard
Terminal