MODEL 6292
SERVICE/CALIBRATION MANUAL
CAT.# 2129.83
Calibration Procedure
Warning:
In calibration mode, the test duration is unlimited. To avoid overheating, finish the
adjustment as fast as possible.
1. Put the AEMC 6292 in calibration mode (see previous procedure).
2. Setup the AEMC 6292 to perform a test using the 1000
μΩ
shunt.
3. Press the START key to initiate the test.
4. Type the character
'r'
through HyperTerminal to see the operating range.
5. Using the current adjust knob, increase the current to 200A, wait a few seconds for the resistance value to
become stable and perform the adjustment sending the appropriate command through HyperTerminal.
6. After reaching the target value, save the adjustment type the character
's'
.
7. Stop the test.
8. Setup the AEMC 6292 to perform a test using the 500m
Ω
shunt.
9.
Repeat the same steps to calibrate the second range of resistance,
but use 10A instead of 200A;
v1 - 03/15