MODEL 6292
SERVICE/CALIBRATION MANUAL
CAT.# 2129.83
9. In the HyperTerminal window, a Password will be requested. Type the character
'k'
(lowercase).
10. The following screen will be displayed.
•
The first line indicates the equipment has entered calibration mode.
•
The second line shows the test setup, which is irrelevant because
in calibration mode the AEMC
6292 will always work manually for an unlimited amount of time.
•
The third line is important to show the equipment is recognizing the main power supply frequency. This
way it will be able to select the appropriate filter to work.
Commands
Character Function
r
Shows the resistance operating range and it’s status.
d
Fine adjustment of resistance value down.
D
Adjustment of resistance value down.
u
Fine adjustment of resistance value up.
U
Adjustment of resistance value up.
s
Saves the last adjustment.
Adjust mode
Normal (200A – 120s)
V=220V f=60Hz
09/03/13 04:27 PM
v1 - 03/15
- 6 -