MODEL 6292 

SERVICE/CALIBRATION MANUAL 

CAT.#  2129.83 

 

2.

 

Calibration Check 

Requirements  

 

  One set of standard test cables 

  Shunt 100mVx200A (50

μΩ)

 

  Shunt 166mVx100A (166.6

μΩ)

 

  Shunt 200mVx200A (1000

μΩ)

 

 

Shunt 5Vx10A (500mΩ)

 

 
Check the different values from the chart below: 
 

FUNCTION 

TESTED 

RANGE 

STANDARD 

VALUE 

TOLERANCE 

 

MIN 

MAX 

200A  

2000 µ

 

 50.0 µ

 

49.5 µ

 

50.5 µ

 

 

 

100 µ

 

99.0 µ

 

101.0 µ

 

 

 

1000 µ

 

990.0 µ

 

1010.0 µ

 

 

 

 

 

 

100A 

2000 µ

 

50 µ

 

49.5 µ

 

50.5 µ

 

 

 

100 µ

 

99.0 µ

 

101.0 µ

 

 

 

166.6 µ

 

165.0 µ

 

168.3 µ

 

 

 

1000 µ

 

990.0 µ

 

1010.0 µ

 

 

 

 

 

 

10A 

 

500 m

 

495.0 m

 

505.0 m

 

Accuracy: 1% of reading

 

 

 

 
 

If one of the values is not correct, follow the calibration procedure in the next section. 
 

v1 - 03/15 

- 4 -

 

Summary of Contents for 2129.83

Page 1: ...MC Model 6292 Cat 2129 83 This document is confidential property of Chauvin Arnoux Inc d b a AEMC Instruments and may not be reproduced without permission AEMC Instruments retains the right to make changes to specifications herein at any time without notice ...

Page 2: ...ation 5 Requirements 5 Entering Calibration Mode 5 Commands 6 Calibration Procedure 7 4 Firmware Update 8 Requirements 8 Firmware Update Procedure 8 5 USB Interface Update 11 Requirements 11 USB Interface Update Procedure 11 Repair and Calibration 12 Technical and Sales Assistance 12 Limited Warranty 13 Warranty Repairs 13 v1 03 15 2 ...

Page 3: ...te through menu Check if clock is running after some minutes LCD Backlight try by menu LCD Contrast try by menu Generation of 200A and resistance measurement Grounding loop Test using the auxiliary clamp in the two senses ways Memory Save a test result Recall a test saved Erase the memory Fan On mode Auto mode Check the high voltage detection on the output bornes using 30VAC Safety generation cutt...

Page 4: ...es from the chart below FUNCTION TESTED RANGE STANDARD VALUE TOLERANCE MIN MAX 200A 2000 µΩ 50 0 µΩ 49 5 µΩ 50 5 µΩ 100 µΩ 99 0 µΩ 101 0 µΩ 1000 µΩ 990 0 µΩ 1010 0 µΩ 100A 2000 µΩ 50 µΩ 49 5 µΩ 50 5 µΩ 100 µΩ 99 0 µΩ 101 0 µΩ 166 6 µΩ 165 0 µΩ 168 3 µΩ 1000 µΩ 990 0 µΩ 1010 0 µΩ 10A 500 mΩ 495 0 mΩ 505 0 mΩ Accuracy 1 of reading If one of the values is not correct follow the calibration procedure ...

Page 5: ...Entering Calibration Mode 1 Connect the USB cable from the computer to the powered off AEMC 6292 2 Launch HyperTerminal 3 Configure HyperTerminal Com Port as shown below same as Control Panel Device Manager 57600 bps 8 bits No Parity No Flow Control 4 Start the connection in HyperTerminal 5 Press and hold the STOP key 6 Power on the instrument while pressing the STOP key 7 When the auto tests comp...

Page 6: ...always work manually for an unlimited amount of time The third line is important to show the equipment is recognizing the main power supply frequency This way it will be able to select the appropriate filter to work Commands Character Function r Shows the resistance operating range and it s status d Fine adjustment of resistance value down D Adjustment of resistance value down u Fine adjustment of...

Page 7: ...tiate the test 4 Type the character r through HyperTerminal to see the operating range 5 Using the current adjust knob increase the current to 200A wait a few seconds for the resistance value to become stable and perform the adjustment sending the appropriate command through HyperTerminal 6 After reaching the target value save the adjustment type the character s 7 Stop the test 8 Setup the AEMC 62...

Page 8: ... Procedure 1 Connect the USB cable from computer to the powered off AEMC 6292 2 Launch HyperTerminal 3 Configure HyperTerminal Com Port as shown below same as Control Panel Device Manager 57600 bps 8 bits No parity No Flow Control 4 Press and hold the ZERO and TWO keys 5 Power on the instrument while pressing the ZERO and TWO keys 6 The following screen will be displayed FIRMWARE UPDATE v1 03 15 8...

Page 9: ...sented by a yellow phone to start the connection 8 If the connection was made successfully the S character will be displayed sequentially in HyperTerminal screen 9 In the main menu choose Transfer Send File 10 Select the new firmware file 11 In the same window choose the Xmodem protocol v1 03 15 9 ...

Page 10: ...UAL CAT 2129 83 12 Press Send to start firmware update The following window will be displayed 13 At the end of the firmware update the window above will automatically be closed and the application will start in the equipment v1 03 15 10 ...

Page 11: ...scription information in the USB module follow the steps below 1 Connect the computer to the AEMC 6292 using an USB cable 2 Launch the Mprog exe 3 In the main menu choose File Open find and select the file AEMC6292 ept located in the Template directory 4 In the main menu choose Device Program to perform the update 5 If successful the message Programmed Serial Number xxxxx will be displayed in the ...

Page 12: ... T includes calibration certificate plus recorded calibration data Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 repair aemc com Or contact your authorized distributor Costs for repair standard calibration and calibration traceable to N I S T are available NOTE All customers must obtain a...

Page 13: ...ne year period you may return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material Warranty Repairs What you must do to return an Instrument for Warranty Repair First request a Customer Service Authorization Number CSA by phone or by fax from our Service Depar...

Page 14: ...over NH 03820 USA Phone 603 749 6434 800 945 2362 Fax 603 742 2346 www aemc com This document is confidential property of Chauvin Arnoux Inc d b a AEMC Instruments and may not be reproduced without permission AEMC Instruments retains the right to make changes to specifications herein at any time without notice ...

Reviews: