MODEL 6292
SERVICE/CALIBRATION MANUAL
CAT.# 2129.83
2.
Calibration Check
Requirements
•
One set of standard test cables
•
Shunt 100mVx200A (50
μΩ)
•
Shunt 166mVx100A (166.6
μΩ)
•
Shunt 200mVx200A (1000
μΩ)
•
Shunt 5Vx10A (500mΩ)
Check the different values from the chart below:
FUNCTION
TESTED
RANGE
STANDARD
VALUE
TOLERANCE
MIN
MAX
200A
2000 µ
Ω
50.0 µ
Ω
49.5 µ
Ω
50.5 µ
Ω
100 µ
Ω
99.0 µ
Ω
101.0 µ
Ω
1000 µ
Ω
990.0 µ
Ω
1010.0 µ
Ω
100A
2000 µ
Ω
50 µ
Ω
49.5 µ
Ω
50.5 µ
Ω
100 µ
Ω
99.0 µ
Ω
101.0 µ
Ω
166.6 µ
Ω
165.0 µ
Ω
168.3 µ
Ω
1000 µ
Ω
990.0 µ
Ω
1010.0 µ
Ω
10A
500 m
Ω
495.0 m
Ω
505.0 m
Ω
Accuracy: 1% of reading
If one of the values is not correct, follow the calibration procedure in the next section.
v1 - 03/15
- 4 -