Normal operation will be impossible for refrigerator if it is installed in 

environments other than those described above as it is sensitive to ambient 

temperature. Start operation after the environment is improved.

Ambient temperature: 10

 to 32

, 18

 to 25

 are optimal and air conditioning system is 

required as necessary.
Ambient humidity: below 85%RH.

Avoid excessive dust.
Avoid mechanical swing or vibration.

The refrigerator shall operate at an altitude lower than 2000m.
Input voltage: no greater than ±10% of rated voltage.

It is prohibited to install the refrigerator outdoor. Electric leakage or shock may 

be caused if the refrigerator gets wet by rainwater.

The installation site shall meet following requirements for normal operation and best performance of 

refrigerator:

Do not install the refrigerator in a narrow and confined space and the door of the installation space 

may not be smaller or lower than the refrigerator, to ensure smooth entry and exit of the refrigerator 

as well as to prevent damage to the stored articles due to not timely repair of refrigerator as the 

installation site is not easy to be accessed.
The ground at the installation site shall be solid and flat.
It is well ventilated and free from direct sunshine.
Do not share one socket with plugs of other equipment and the plug shall be securely plugged into 

the socket.

Do not twist or compress the power line.
If the power line needs to be extended, the extended line shall be no smaller than 3mm² in 

sectional area and no longer than 3m.
Check the operating voltage before operation and voltage regulator appropriate for motor load may 

be applied in regions with unstable voltage.Ensure the normal input voltage stands at rated 

voltage ± 10% and the power of voltage regulator is greater than 4000W.
The refrigerator shall be grounded in a reliable manner.
If the socket is provided with grounding wire, check if it is well grounded before operation.
If the socket is not provided with grounding wire, a new one shall be installed by professional 

engineers.

Installation environment

Installation site

Do not ground the refrigerator through gas pipes, water supply pipes, telephone 

lines or lightning conductors as electric shock may be caused.
The power plug can be accessed after installation to ensure the power line can 

be pulled out in a timely manner in emergency cases. The air vent shall free from 

barriers.

Product Installation

Battery maintenance

When the refrigerator works continuously, please test battery charge level every 15 days (please 
refer to “Alarm Test” on Page 18 for test method). In case of low battery charge level, make sure 
that battery switch is on and battery is charged. When the battery has been charged for consecutive 
7 days, please retest battery charge level. In normal conditions, battery charge should be full. If low 
battery charge level remains, it is recommended to replace the rechargeable battery.The battery of 
power failure alarm is a consumable item with 3-year service life. If it has been used for more than 3 
years, power failure alarm may not act when alarm conditions occur. It is recommended to replace 
the battery earlier. Please contact a Haier after-sales service technician for battery replacement.

b.Battery removal
1)Pull out the plug from the power socket.
2)Screw off screws on the cover of electric cabinet with a screwdriver.
3)Pull out the battery's connecting plug.
4)Remove the battery's fixing clip and take out the battery.
5)Recycle or properly dispose the battery as required.

When replacing the battery, make sure that red wire is connected to the battery's positive pole 
and black wire to the battery's negative pole. Do not misconnect otherwise the charging circuit of 
computer board would be burned out easily and thus the battery cannot be charged.

The refrigerator has a built-in rechargeable battery. The battery is recyclable; when its service life 
expires,  please ask a local relevant battery recycling organization for check or discard it properly.
a.Battery location
The built-in battery of the refrigerator-freezer is used for power failure alarm and located in electric
cabinet.

There are high-voltage electrical components in electric cabinet.
To avoid electric shock, the cover of electric cabinet must be opened by a 
qualified engineer or maintener only.

Rechargeable battery recycling

Prohibition

Summary of Contents for EL-RG-49H

Page 1: ...or may vary Translation of the original instruction The scope of the Product Application Pharmaceuticals and other related products which need a storage environment of 2 8 Certi cate of Quality Pharma...

Page 2: ...harmaceutical Refrigerator Type Designation Trademark EL RG 49H Manufactures Name Aegis Scientific Inc Address Tel 1 Ivybrook Blvd Unit 145R Ivyland PA 18974 USA 1 800 796 2344 Fax 732 879 0248 Refer...

Page 3: ...essor and main components and better refrigeration performance User friendly design computer control smart and carefree adjust ment not required High performance thermal insulation layer with excellen...

Page 4: ...the appliance that you have just purchased due to technical improvements 1 Product Feature Content Safety Precautions Precautions for Use Product Installation Component Names Control Panel Applicatio...

Page 5: ...ug in or pull out the power plug of refrigerator to prevent explosion and fire The refrigerator shall be installed by professional technical staffs or after sale maintenance staffs to prevent electric...

Page 6: ...ay pose danger to human health or environment Pull out the power plug if the refrigerator has been out of service for long time to prevent electric shock leakage or fire caused by aging power lines If...

Page 7: ...ust be placed on a solid and flat surface or excessive vibration and noise may be produced when the appliance in operation The appliance can be used by children aged from 8 years and above and persons...

Page 8: ...hazardous substances can leak into the groundwater and get into the food chain damaging your health and well being When replacing old appliances with new ones the retailer is legally obligated to tak...

Page 9: ...ion site Do not ground the refrigerator through gas pipes water supply pipes telephone lines or lightning conductors as electric shock may be caused The power plug can be accessed after installation t...

Page 10: ...ownward upward Clean the back evaporating box Evaporating box is installed on the back of the refrigerator Remove 2 screws on the bottom loosen the upper 2 screws no need to remove them lift the evapo...

Page 11: ...m is under normal condition When chargeable battery is closed or with too low a power press once the alarm testing key E6 is expected to flash 3 times on the display window at the simultaneously The r...

Page 12: ...220V 110V failed under any of the above conditions remote alarm shall be started Coupled with 1Hz continuous beeping Take corres ponding measures according to item 1 2 and 3 Remote alarm startup when...

Page 13: ...upper or lower temperature displayed ALH Alarm Limit High settable by the user high temperature alarming is triggered while when any of upper or lower temperature shown ALL Aalrm LImit Low settable by...

Page 14: ...inside temperature display area flashes Date press Minus button or Plus button to adjust the current date and the press Set button to save and confirm Thereafter the inside temperature display area sh...

Page 15: ...e display area steadily shows 2P press Set button the inside temperature display area flashes Month factory default 01 press Minus button or Plus button select Month from 01 12 recycling display after...

Page 16: ...arameter value After setting press Set to go back and choose other parameters for the same token With all values set press Set button 5s exit and return to the normal display mode Temperature Display...