--------------------------------------------------------------------------------25/28--------------------------------------------------------------------------------
Figure 10: Software triggered reading operation with TOR>0
Please note
: The TOR parameter is only active, if the yT-Command is applied. Within the time span
defined by the value of TOR no NoRead will be output on the interface!
4.2
MD 0 - continuous reading
When operating continuously the exciter is switched on permanently. The reading cycles are initiated
periodically.
After an accomplished reading cycle the reading information is evaluated. After that data (either tran-
sponder number or NoRead code) is output to the serial interface
Figure 11: continuous operation
exciter
processor
interface
TOR
GT
reading process
NoRead
reading cycle reading cycle
reading cycle
exciter
processor
interface
reading cycle
ID
ID
ID
reading cycle
reading cycle