26
5.
Step-by-Step Software Use
The following sections will explain, step-by-step, how to initialize, configure, set-up, test,
view data, and all other software-driven functions of the WTD Model 501.
5.1
OVERVIEW OF USE
After setting up the hardware, and placing the hand-held probe for testing, the user is ready
to begin using the software. To setup the WTD: see Section 3.
To perform the wedge tightness testing, the WTD needs examples of a tight wedge and a
loose wedge in order to set the appropriate data analysis and processing parameters. These
reference values are stored in a calibration file and may be selected or created. See Sections
3.3 and 5.4 for more details.
The first step before testing the wedges for tightness is either to create or open a unit, or
create a calibration file.
5.1.1
CREATING A UNIT
To create a unit, click on the
New Unit
icon or choose
New
from the
Unit
menu. A dialog
box will appear with three tabs for information to be filled in regarding the unit in which you
will be testing the wedges.
Figure 22: WTD Properties General Tab
The first tab is titled
General
.
Station name, unit name and number of slots and wedges
on the second tab is the minimum required.
The rest of the information on second and
third tabs is optional.