Advanced Telemetry Systems, Inc. R4500S Reference User Manual R05-11-A
17
Screen
3
The number entered here should represent the range of pulse rates you are using in your study. The available
period range allowed to be entered is between 400ms to 6000ms. The periods measured from signals that are
outside the entered range will be thrown out.
2.2.2 Stationary Defaults
Stationary Defaults – Fixed Pulse Rate
Move the cursor to the Stationary Default item and press
“ENTER”
. There will be multiple windows to go through.
To change the numeric value, use the numeric digits on the keypad, or you can use the
and
keys to
increment the numeric value, up or down or to toggle between YES / NO selections. For the R4500S to accept a
value, press
“ENTER”
.
Moving the cursor with the
or the
will allow movement between screens or by pressing
“ENTER”
for the
last item in the window to proceed to the next window.
Screen
1
Continual:
The R4500S will continually scan through the selected frequency table.
Periodic:
The R4500S will scan through the selected frequency table only once and wait to
scan the frequency table once again until an entered amount of time has elapsed.
The amount of time entered between scans (Log Period) is entered when starting
a Stationary Scan.
Screen 2
The number entered here should represent the total number of pulse rates you are using in your study. There
should not be more than one transmitter per frequency. If you need to use multiple transmitters per frequency you
should consider using a coded receiver and coded transmitters.