Calibration
•
69
6.4 D/A Adjustment
6.4.1 Unipolar Analog Output
1. Set JP1 to select unipolar. Connect VDM (+) to CN3 pin-30 (DAOut) and
VDM (-) to A.GND.
2. Write the digital value 0 to DAC. T rim
VR2
to obtain 0V reading in the DVM
3. Write the digital value 4095 to DAC. Trim
VR1
to obtain 10V reading in the
DVM.
6.4.2 Bipolar Analog Output
1. Set JP1 to select bipolar. Connect DVM (+) to CN3 pin-30 (DAOut) and
DVM(-) to A.GND.
2. Write the digital value 2048 to DAC. Trim
VR2
to obtain 0V reading in the
DVM.
3. Write the digital value 4095 to DAC. Trim
VR1
to 10V reading in
the DVM.
A calibration utility is supported in the software CD which is included in the
product package. The detailed calibration procedures and description can be
found in the utility. Users only need to run the software calibration utility and
follow the procedures. You will get the accurate measure data.
In normal condition, the PCI-9111 already calibrated by factor before it is
shipped out. So, users do not need to calibrate your PCI-9111 when you get it.