Operation Theorem
•
35
4.4 D/A Conversion
The PCI-9111 has one analog output channel. The signal range can be
uni-polar or bi-polar which are set by JP1.
To D/A Output
D/A Converter
-
+
-10V
Ref In
Analog GND
Pin-30 (DA Out)
The operation of D/A conversion is simpler than A/D operation. You only need
to write digital values into the D/A data registers and the corresponding voltage
will be output from the DA Out (pin-30 of CN3). Refer to section 3.15 for
information about the D/A data registers. The mathematical relationship
between the digital data
DAn
and the output voltage is formulated as following:
V
out
=
span x
DAn
/ 4096
– Unipolar
V
out
=
span x
DAn
/ 4096 + (-10)
– Bipolar
where
span
is the span in volts. If your output range is -10V~10V (Bipolar),
then span is 20; if your output range is 0~10V (Unipolar), then span is 10. The
V
out
is the output voltage, and the
DAn
is the digital data value in the D/A data
registers.
Before performing the D/A conversion, users should care about the D/A output
range which is set by the JP1. Please refer section 2.4 for jumper setting.
Analog Output
Digital Data Input
Unipolar
0V ~ 10V
Bipolar
-10V ~ 10V
FFF hex
+9.9976V
+9.9951V
800 hex
+5.0000V
0.0000V
7FF hex
+4.9976V
-0.0049V
000 hex
0.0000V
-10.0000V
1 LSB
2.44mV
4.88mV