Appendix B
Specifications
29
B
A P P E N D I X
Cycle Operation
Max Inflation Pressure:
200 or 280 mmHg (switch-selectable)
Measurement Cycle Time:
40–150 BPM (large animals): 44 s (for 200 mmHg)
87 s (for 280 mmHg).
90–420 BPM (rats):
22 s (for 200 mmHg)
41 s (for 280 mmHg).
240–600 BPM (mice):
18 s (for 200 mmHg)
40 s (for 280 mmHg).
Cycle Control:
Microprocessor based. Performs inflation,
deflation, and fast deflation sequences
automatically.
Operation Indication:
Trigger output (normally low, but high during
inflation and deflation cycle).
Front panel Status indicator shows inflation
and deflation operation.
Operation Abort:
The cycle can be terminated at any point by
pressing the front panel Start/Stop button
again. Pressure is automatically released.
Fast Release Time:
~1.2 s from 280 to 40 mmHg at 40–150 BPM
~0.5 s from 280 to 40 mmHg at higher ranges.
Control Sources:
Front panel push button; External signal
source (voltage level); Remote contact closure.
B
Specifications