NIBP Controller
Owner’s Guide
30
Manual Start/Stop Input
Operation:
Contact the closure input for starting or
stopping measurement cycle. Shorting the
input signal results in a start or stop operation.
Minimum Contact Closure:
1 ms
Remote Trigger Input
Operation:
Voltage level input for starting or stopping
NIBP cycle. TTL compatible input. High level
operates a start/stop.
Input Voltage:
3–5 V
Minimum Trigger Pulse:
1 ms
Trigger Output
Operation:
High (+5V) output level during measurement
cycle. Otherwise zero.
Pressure Output (Cuff Pressure)
Sensitivity:
0–1 V : 0–300 mmHg (factory calibrated)
Frequency Response:
DC to 10 Hz
Pulse Input
Input Impedance:
Differential Input: 10 G
Ω
Single-ended Input: 1 M
Ω
Input Signal Range:
High gain differential input: (0–25 μV) up to
(0–75 μV), depending on rear panel Gain Adj
setting.
Low gain single-ended input: (0–50 mV) up to
(0–150 mV), depending on rear panel Gain
Adj setting.
Bandwidth:
40–150 BPM:0.7–2.5 Hz
90–420 BPM:3–7 Hz
240–600 BPM:4–10 Hz
Pulse Output Max:
± 5 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com