![ADE STAN07 Operating Manual Download Page 27](http://html1.mh-extra.com/html/ade/stan07/stan07_operating-manual_2849736027.webp)
STAN07-210324-Rev005-UM-en
26
Confirm your selection with
to change to the next menu item or press the
key to return to the
previous menu.
Data output via serial port (second display or PC)
When set to C1 (for secondary display) or C2 (for PC), data output takes place via the serial interface
in different data structures.
Data output when selecting C1 (Continuous 1) for secondary display
Data output after choosing C1 in the data structure:
Status
Indicated
Tare
STX STA
STB
STC
D5
D4
D3
D2
D1
D0
D5
D4
D3
D2
D1
D0
CR
LF
CHK
Referring to the status bytes STA, STB and STC:
Definition Overview for Status A (STA)
Definition Overview for status B (STB)
Bits 0, 1 and 2
Bits 3 and 4
Bits 5,6 Bit 7
Bit 0
Bit 1
Bit 2
Bit 3
Bits 4,5
Bit 6
Bit 7
0
1
2
Decimal
point
3
4
Inc.
size
A
lw
a
y
s
1
X
1
=
N
e
t
1
=
W
e
ig
h
t
n
e
g
a
ti
v
e
1
=
E
rr
o
r
1
=
U
n
s
ta
b
le
A
lw
a
y
s
=
1
1
=
Z
e
ro
e
d
w
it
h
p
o
w
e
r
o
n
z
e
ro
X
1
0
0 XXXXXO
1
0 X 1
0
1
0 XXXXXX
0
1 X 2
1
1
0 XXXXX.X
1
1 X 5
0
=
G
ro
s
s
0
=
W
e
ig
h
t
p
o
s
it
iv
e
0
=
N
o
E
rr
o
r
0
=
S
ta
b
le
0
=
N
o
t
p
o
w
e
r
o
n
z
e
ro
e
d
0
0
1 XXXX.XX
1
0
1 XXX.XXX
0
1
1 XX.XXXX
Status C (STC) is always hex '30'.
Note: The weight data are shown aligned to the right in the display, error messages (under, over, A.Out)
to the left.
Data output when selecting C2 (Continuous 2) for PC
Data output after choosing C2 in the data structure:
[STX] [STATUS] [SIGN] [INDICATED WEIGHT VALUE] [UNIT] [CR] [LF] [CHK]
Examples:
S+000123.4kg
(Weight is stable and 123.4)
D+000123.4kg
(Weight is dynamic and 123.4)
+
(Overload)
-
(Under load)
O
(ADC out error)
Checksum
CHK (Checksum) = 0 – (STX
STATUS
.....
LF)
Printer settings
In this menu, the print format, line feed, printing language, number of printouts, and the printer type can
be specified.
To make changes, first press the
key when the display reads [dAtA] and ► many times until [Print]
appears in the display. Confirm with the
key.
Summary of Contents for STAN07